Gate: Used for analyzing and measuring the pulse.
The Gate function is useful for setting the location and area of defects and also for checking thickness differences.
You can also configure the Gate by clicking the icon on the right

Gate: A feature that can be utilized in various ways such as locating defects, collecting data range, and setting references.
Gate: Consists of three gates (A, B, I).
Gate Start: Sets the start position of the selected gate.
Gate Range: Sets the range of the selected gate.
Gate Threshold: Sets the height of the selected gate.
Add / Remove: Allows you to add or remove the selected gate.
Applied based on the selected gate.
Echo Type: A menu to set the measurement format of the gate.

Sound Path: An indicator and function that helps measure the distance to defects based on the beam.
True Depth: An indicator and function that helps measure the distance to defects based on the material's thickness.
Detection: A menu for setting the criteria when measuring signals with the Gate. (Divided into two types)

Peak: The location with the highest amplitude among the signals within the gate range.
Edge: The point where the gate meets the signal within the gate range.