Wedge Delay: A function used for distance calibration for the entire angle (vector) set in the Focal Law.

Based on the first reflected signal, input the Ref value. The red line should be inside the green line.


  1. Ref: A menu to enter the distance value of the signal to be calibrated.

  2. Tolerance: A menu to set the error range, which corresponds to the green line.

  3. Clear Envelope: Used to reconfigure the red line in the calibration window.

  4. Reset: Resets the previously set value.

  5. Apply: A button selected when the red line enters the green line.


  1. Finish: Ends WD Calibration.

    - After the calibration is complete, a W will light up at the bottom.