top of page


 

TCG: A calibration function similar to Sensitivity.

The difference from Sensitivity is that it allows calibration across multiple ranges.

<TCG Display>
<TCG Display>

※TCG should be calibrated sequentially from the shallow depth area to the deeper depth area.


<TCG Calibration before Display>
<TCG Calibration before Display>

  1. Ref: Enter the reference value for the amplitude.

  2. Tolerance: Input the allowable error to specify the range (applied to the green range).

  3. Step: Indicates the current position being calibrated (#1, #2, …).

  4. Step mm: A menu to set the desired movement distance when manipulating the arrows.

  5. ◀ / ▶: A menu to adjust the position of the step.

  6. X (Delete): A menu used to delete the step.

  7. Clear Envelope: Reconfigure the red line in the calibration window.

  8. Angle Start/Stop: Set the angle range for calibration.

  9. Reset: Reset the set TCG calibration value.

  10. Apply: Click this button to apply the TCG calibration value.

  11. Finish: End TCG calibration.

    - After the calibration is complete, a T will light up at the bottom.

<TCG Display>
<TCG Display>
<TCG Calibration after Display>
<TCG Calibration after Display>