Sub Toolbar: This list is necessary for using measurement-related functions on the screen.
Sub Toolbar: It can be activated by clicking on the Window list in the Total Menu.
Parameter

Parameter: This screen is used to check the values stored during data collection.
※ (Depth, Pulse, Voltage, etc…)
Colormap

Color Map: There are two types of color maps.
The first type is the commonly used Color Map, based on A^.
The second type is the Color Map used for Corrosion processes.
The Corrosion Color Map is automatically applied when changing to A/, B/, A/-B/, Position.
※ To further distinguish color differences, there is a function that allows you to reset the range differences using Min/Max. The colors can also be smoothed or reversed using the Linear and Reverse functions.
Measure Tool
Measure Tool: A list that provides position value information for Gate, Bar, and Laser.

Gate
Composed of A / B / I, the following provides explanations for each of the values.
Button | Unit | Explanation |
A%, B%, I% | % | Amplitude (%) Screen Height for A, B, and I. |
SA, SB, SI | mm | Soundpath of Echo in A, B, and I Gates |
DA, DB, DI | mm | Reflector depth of A, B, and I Gates |
PA, PB, PI | mm | Projection distance of echo in A, B, and I Gates |

Detection
This section sets the method for reading waveforms that pass through the gate.
※ Peak: This function reads based on the highest point of the waveform within the gate.

※ Edge: This function reads based on the point where the waveform touches the gate.

Line Bar
Line Bar: A list of information that reads the position values of Ref, Meas, and Data. M-R shows the calculated difference between Ref and Meas.
Gate/Range Control
Gate / Range Control: A list that allows you to reset the position settings of the gate and the settings for View Range / Calibration.
Gate: You can click on the desired Gate through Kind to set Start / Range / Threshold and change the position.
View Range: This function is frequently used when you want to view only the desired area, excluding other positions. If you have roughly set the position in the Scan window, you can use the View Range Control list to adjust the fine position values.
Calibration: This function shows the Calibration values performed during data collection and is used when you need to modify the set WD / Velocity values. (Note: Applies only to Ruler.)